Handheld Air Particle Counter Y09-3016E
This product adopts a color LCD touch screen and is the first handheld dustparticle counter in China to adopt this design.
1. Conforming to the verification specifications of JJF-1190-2008dustparticle counter
2. Self purification time1Omin
3. Measurement repetition relative deviation10%.
4. Particle size distribution error 0.5um 5um20%.
5. Particle concentration indication error 0.5um 20%.
6. Sampling flow rate and error: 2.83L/min 5%
7. Measurable particle size range: 0.3,0.5,1,3,5,10um, divided into6 particle
size channels, with a signal-to-noise ratio of 2.5:1 for 0.3um.
8. Maximum measurable particle concentration:30Kparticles/L.
9.Current DC7.2V, 1800mAh.
Specification
Outside Dim |
300*360*126 mm3(Wide*Deep*High) |
| Display |
LCD |
| Weight |
0.6kg |
| Battery |
(Lithium Ion Battery),7.2V,2200mAh |
| Power Source |
Li-ion charger ,8.4V DC,1A:9-250V AC, 50-60Hz |
| Size Channels |
0.3, 0.5, 1, 3, 5, 10 μm |
| Flow Rate |
2.83L/min |
| Sample Time |
1s-99h |
| Cycles |
0-99 |
| Hold Time |
0-999sec |
| Delay Time |
0-99sec |
| Light source |
Laser Diode |
| Zero Court |
<=10min |
| Alarm |
Low Battery ;100, 1000, 10000 Class /ISO |
| Coincidence Loss |
5%,2000000ps/cf |
| Operating Time |
5 hours |
| output |
RS-232 to pc or printer via optional cable |
| Location Labels |
0-999 |
| Data Storage |
1000 Samples (With PC Software) |
| Environmental |
10degrees celsius-40degrees celsius ; Humidity: 20%-90% RH ;Atmosphere Pressure 86Kpa-106Kpa |
| Standard feathure |
suitcase ;charger;sampling head;line of communication;communication software ;Automatic Purifier |
| Optional component |
External printer
|